FILTRATION AND VACUUM DESICCATORS
DescriptionImplosion proof desiccators. Will take vacuum of -740 mm Hg with loss (absolut zero: -760 mm/HG) not exceeding 20 mm Hg in 24 hours. The transparent polycarbonate cover seals onto a neoprene ‘0’ ring set into the flange of the PP base. A little grease spread on the ‘0’ ring will assist in establishing initial vacuum. Within the desiccator is a removable pan for holding a drying agent. Vacuum release tap/non return valve assembly is included (Art. 229); specially grooved stopper in this assembly allows air, when required, to be re-admitted very slowly to avoid disturbing the contents. Usage Instructions:
• remove the cover in PC - insert the samples in the pan
• open the vacuum retention valve (PE needle) connect the tube of the pump to the dessiccator spigot in order to create vacuum
• close the vacuum retention valve
Art. | Vol. lt | Int. Ø mm | Ext. h mm | Int. h mm |
---|---|---|---|---|
550 | 2,15 | 150 | 190 | 135 |
230 | 4,35 | 200 | 230 | 175 |
554 | 9,20 | 250 | 300 | 225 |